Viscoelastic Mattress MT 02
195 x 85 cm patient mattress with a viscoelastic top layer and a 28 kg/m³ HD polyurethane base layer; air channels prevent heat build-up, and the…
Dynamic alternating mattress with 20 TPU air cells and a dual-compressor digital pump for patients at medium to very high pressure ulcer risk; recommended maximum weight 255 kg.
The Schröder MT 07 is an air-cell dynamic mattress system aimed at preventing pressure ulcers. Its high-quality TPU-coated cells provide reliable and effective therapeutic support. The mattress has 20 dynamic air cells working in a 2-1 alternating pattern and 3 static cells in the head section; the 20 cm high cells inflate and deflate in turn with the digital pump. This design suits patients whose risk of developing pressure ulcers ranges from medium to very high.
Programmed by a microcontroller, the multi-function digital pump runs on two linear compressors with a flow rate of 9.0 L + 9.0 L and has a detachable power cable. Its high flow rate lets the mattress inflate quickly. A precise digital pressure transducer lowers the power drawn once the target pressure is reached, which is what makes the pump quiet, energy-efficient and long-lasting. The pump offers a 20-minute static mode, an alternating mode with selectable cycle time, a constant low pressure (CLP) or pulsating mode, and auto-firm and auto-sensing modes. When pressure drops, the power fails or a fault requiring service occurs, the pump gives an audible and visual alarm and shows the fault as a code; these warnings protect patient safety.
| Mattress Dimensions | 200 x 90 x 20 cm (8-inch replacement type) |
|---|---|
| Number of Cells / Material | 20 / TPU |
| Top Cover Material | Two-way stretch polyester |
| Head Cushion Section | Yes, 3 cells |
| Independent Low Air Loss (iLAL) | Yes (up to 75 L per cycle) |
| CPR Type | AKTC Q-CPR |
| Recommended Maximum Weight | 255 kg |
| Pump Type / Size | Digital / 43 x 13 x 24 cm |
| Flow Rate | 9.0 + 9.0 L |
| Supply Voltage | 230 V / 50 Hz |
| Modes | Alternating, static, constant low pressure (CLP) |
| Full Cycle Time | Selectable: 10, 15, 20, 25, 30 minutes |
| Classification | Class II, Type BF |
| Electrical Safety Standards | IEC 60601-1, IEC 60601-1-2, IEC 60601-1-11 (home care) |
For every device you purchase we set up an installation, training and after-sales support plan. Contact our teams for preventive maintenance contracts, spare parts and clinical application support.
195 x 85 cm patient mattress with a viscoelastic top layer and a 28 kg/m³ HD polyurethane base layer; air channels prevent heat build-up, and the…
Hospital mattress made of 28 kg/m³ PU foam with CNC-cut surface and air channels. Supplied with an antibacterial, waterproof, breathable PU cover…
Two-layer anti-decubitus mattress with a CNC-cut cubic soft foam top and a high-density PU foam base. Regionally varied cut sizes aim to reduce…
Comfort-focused hospital mattress with a soft foam top and a 28 kg/m³ high-density PU foam base. Tube-design air channels support air circulation.
Two-layer mattress for patients at high risk of decubitus ulcers: CNC-cut viscoelastic foam on top, high-density PU foam below; standard size 85 x…
Three-layer hospital mattress, 12 cm thick, built from viscoelastic, foam and cold foam layers. Designed for the comfort of patients at high risk of…