Discharge Valve
Brass valve designed for discharging gas from medical gas stations. Compatible with O2 and N2O.
Area valve service module for isolating the medical gas lines on each hospital floor separately. Meets HTM 02-01 and EN ISO 7396-1 requirements; 3, 4 or 5 gases, with or without alarm.
The AVSU (Area Valve Service Unit) module is produced so that the medical gas lines in a hospital can be isolated separately on each floor. During installation and maintenance work or in emergencies, the relevant line may need to be taken out of service; the module makes this isolation possible.
The unit has all the features required by the HTM 02-01 and EN ISO 7396-1 standards. It can be used on oxygen, nitrous oxide, 400 kPa medical air, 700 kPa surgical air and medical vacuum lines, and is available in under-plaster and on-plaster versions, with or without alarm.
| Standards | EN ISO 7396-1, HTM 02-01 |
|---|---|
| Services | Oxygen, nitrous oxide, medical air 400 kPa, surgical air 700 kPa, medical vacuum |
| 3 Gas, Without Alarm | Under plaster: AVSU-30; on plaster: AVSU-31 |
| 3 Gas, With Alarm | Under plaster: AVSU-30A; on plaster: AVSU-31A |
| 4 Gas, Without Alarm | Under plaster: AVSU-40; on plaster: AVSU-41 |
| 4 Gas, With Alarm | Under plaster: AVSU-40A; on plaster: AVSU-41A |
| 5 Gas, Without Alarm | Under plaster: AVSU-50; on plaster: AVSU-51 |
| 5 Gas, With Alarm | Under plaster: AVSU-50A; on plaster: AVSU-51A |
For every device you purchase we set up an installation, training and after-sales support plan. Contact our teams for preventive maintenance contracts, spare parts and clinical application support.
Brass valve designed for discharging gas from medical gas stations. Compatible with O2 and N2O.
Chain that prevents gas cylinders from tipping over and keeps them stable and secure on the ramp. Alternative models for 1 to 5 cylinders.
Ramp that gathers gas cylinders onto one line to feed the system with backup and uninterrupted supply. Galvanized steel, brass heads and copper pipe…
Automatic changeover manifold feeding oxygen, nitrous oxide and other medical gases without interruption; it switches to the reserve cylinders when…
Panel that gives audible and visual warnings when a pressure problem occurs in central gas station manifolds; it monitors pressure via one positive…
Ball valve that controls medical gas flow and quickly shuts off the gas to a given area during emergencies and maintenance. Available with or without…
Zone control unit that manages medical gas lines section by section and provides precise pressure regulation at the point of use. Flush and surface…
140 cm tail pipes and 60 cm flexible hoses for connecting gas cylinders to a cylinder ramp. Available with different fittings for O2, CO2 and N2O.
Flexible connector used to link gas cylinders to a cylinder ramp. Supplied with dedicated fittings for O2, N2O, CO2 and other medical gas cylinders.
Digital alarm panels that continuously monitor medical gas source equipment, pipeline pressures and intensive care areas. Models are available from…